AIM-M100 Medical Insulation Monitor

● Insulation monitoring for medical IT systems

● Load and temperature monitoring of isolation transformer
● Device disconnection monitoring
● LED indicators: On, Insulation, Overload, Overtemp
● Alarm relay output
● Test button
●  Insulation fault location
● RS485 (MODBUS-RTU) + CAN
  • product detail
  • parameter
  • ISOLATION POWER CABINET

General

AIM-M100 auxiliary power supply is AC220V, the product does not have insulation fault location. AIM-M200 auxiliary power is DC 24V. It has all the monitoring functions of AIM-M10 and AIM-M100 products, it also supports the insulation fault location. If you need insulation fault location, please choose AIM-M200. The positioning function can also be turned off through the menu.



Function

1) Functions of real-time monitoring and fault alarming of the ground insulation resistance, transformer load current and transformer winding temperature of the monitored IT system;

2) Can be used with insulation fault locator, remotely starting fault-locating and displaying locating results when there are insulation faults;

3) Real-time monitor the line disconnection fault, temperature sensor disconnection fault and the functional grounding line disconnection fault of the monitored system, and give the alarm indication when the fault occurs;

4) Relay alarm output, LED alarm indication and other faults indication functions;

5) Two kinds of fieldbus communication technology, which are used for centralized alarm and display instrument, test signal generator, insulation fault locator and upper computer management software communications, and can monitor the operation status of IT system in real time.

6) Events logging function, which can record alarm occurrence time and fault type and is convenient for operation personnel to analyze the operation conditions of system and promptly eliminate the faults;


Specification


Product Selection


Dimension



Typical Connection


Wiring & Installation

Installation size of AKH-0.66P26 is shown in the following figure:

  • Front View
  • Bottom View


LED Indicator Instruction

Indicator light

Instruction

On

When the device is in normal operation, the indicator flashes, and the flickering frequency is about once a second.

Insulation

When the insulation resistance exceeds the alarm setting value, or when FK/LL is disconnected, the indicator flashes to alarm.

Overload

When the system load current exceeds the alarm setting value, the indicator flashes to alarm.

Overtemp

When the transformer temperature exceeds the alarm setting value or the temperature sensor wiring is disconnected, the indicator flashes to alarm.


Data Settings

First Menu

Second Menu

Range and Description

SET Comm

Set Addr1

1-247

Set Baud1

4800,9600,19200

Set Addr2

1-247

Set Baud2

4800,9600,19200

Set Alarm

Temp

0-200

Resistance

50-999

Current

14,18,22,28,35,45

Relay Mode

DO-1

Normal Off, Normal On

DO -2

22-24 23;

Set Time

Date

Year-month-day

Time

Hour: minute: second

Set Password

0000-9999kWh

Interface Mode

0-2

Factory Reset

Restore factory settings

Software Version

Firmware version information


Common Problems

Make sure the connection is correct before the system is powered on. For common problems, according to the fault phenomena in the table below, targeted investigation can be carried out.

Fault Phenomenon

Possible Causes and Troubleshooting

LCD: LL break, Insulation indicator lighting.

Terminal 4 and terminal 5 of AIM-M100 are not reliably connected to L1 and L2 of the IT system.

Check the connection and make sure it is reliable.

LCD: FK break, Insulation indicator lighting.

Terminal 13 and terminal 14 of AIM-M100 are not reliably connected to LEB.

Check the connection and make sure it is reliable.

LCD: TC break, Overtemp indicator lighting.

Terminal 11 and terminal 12 of AIM-M100 are not reliably connected to the two ST terminals of the isolation transformer.

Check the connection and make sure it is correct.

LCD: insulation fault, Insulation indicator lighting.

At least one of the L1 and L2 wires in IT system has grounding fault, which can be restored to normal after removal.

The device can't light up.

The working power supply of AIM-M100 is not connected properly.

Check the connection of terminals 1 and 2 and make sure it is correct.


Note: All the above faults should be checked by power off, and the wiring should be adjusted until everything is normal.

AUX Power

Voltage

AC220V±10%

Temperature monitoring

Thermal resistor

2 Pt100

Power consumption

8VA

Measuring range

-50~+200℃

Insulation Monitoring

Resistance measuring range

10~999kΩ

Alarm value range

0~+200℃

Response value

50~999kΩ

Alarm Output

Output mode

2 Relays

Relative uncertainty

±10%

Contact rating

AC 250V/3A

DC 30V/3A

Response time

3s

Environment

Operating temperature

-10~+55℃


Permissible system leakage capacitance Ce

≤5uF

Transport temperature

-25~+70℃

Measuring voltage Um

12V

Storage temperature

-25~+70℃

Measuring current Im

50uA

Relative humidity

5%-95%,No condensation

Impedance Zi

≥200

Altitude

≤2500m

Internal DC resistance Ri

≥240

IP degree

IP30

Permissible extraneous DC voltage Ufg

≤DC280V

Rated impulse voltage / pollution degree

4KV/Ⅲ

Load Current Monitoring

Measuring Value

2.1~50A

EMC

IEC 61326-2-4

Alarm Value

5~50A

Communication

RS485(Modbus-RTU)

Measuring accuracy

±5%


Feedback form contact with us
Related Product
Ground Fault Detection And Location Relay
Ground Fault Detection And Location Relay
Quality Ground Fault Detection And Location Relay manufacturers & exporter - buy Ground Fault De...
Learn More
ASG150 Test Signal Generator
ASG150 Test Signal Generator
You can buy good quality ASG150 Test Signal Generator, we are the best distributor from China market...
Learn More
ACLP10-24 DC Power Supply
ACLP10-24 DC Power Supply
Acrel is the best ACLP10-24 DC Power Supply manufacturer in China. Offering quality product to globa...
Learn More
Product Catalog